Probable edge defect in Acrysof single-piece intraocular lens
Probable edge defect in Acrysof single-piece intraocular lens
Blog Article
The purpose of this article is to report an edge defect in single-piece hydrophobic acrylic intraocular lens (IOL) observed during a routine phacoemulsification procedure.The chip was successfully removed intraoperatively with a McPherson′s forceps.However, six months postoperatively patient complained of a disturbing Flat Iron glare due to posterior capsular folds originating from the site of IOL defect.Patient was advised, Laundry Detergents however refused to undergo Nd:Yag laser capsulotomy to relieve the symptoms.
This case highlights the necessity to check each foldable IOL under the microscope before loading it into the cartridge, for possible edge defects.